Please use this identifier to cite or link to this item: https://hdl.handle.net/10316/4519
Title: Gas proportional scintillation counters with a CsI-covered microstrip plate UV photosensor for high-resolution X-ray spectrometry
Authors: Veloso, J. F. C. A. 
Santos, J. M. F. dos 
Conde, C. A. N. 
Keywords: Gas proportional scintillation counter; CSI photocathode
Issue Date: 2001
Citation: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 457:1-2 (2001) 253-261
Abstract: A xenon-filled gas proportional scintillation counter for X-ray spectrometry is described. The detector uses a microstrip plate covered with a thin CsI film as a photosensor rather than the usual photomultiplier tube. The achieved energy resolution for 5.9 keV X-rays is 12% when the microstrip is operated within the xenon volume and 10.5% when the microstrip is operated in a P-10 environment, separated from the xenon volume by a quartz window.
URI: https://hdl.handle.net/10316/4519
DOI: 10.1016/S0168-9002(00)00746-4
Rights: openAccess
Appears in Collections:FCTUC Física - Artigos em Revistas Internacionais