    The recent development of diffraction-unlimited far-field fluorescence microscopy has overcome the classical resolution limit of similar to 250 nm of conventional light microscopy by about a factor of ten. The improved resolution, however, is susceptible to sample drift on a much finer scale than previously relevant. Here we discuss an approach that corrects for three-dimensional (3D) drift in images of fixed samples without the requirement for fiduciary markers or instrument modification. By imaging mitochondria with Biplane FPALM we show our algorithm’s feasibility in a practical application