Tribocorrosion behaviour of hot pressed CoCrMo−HAP biocomposites
Carregando...
Arquivos
Data
2015-11-01
Autores
Doni, Z.
Alves, A. C.
Toptan, F.
Rocha, L. A. [UNESP]
Buciumeanu, M.
Palaghian, L.
Silva, F. S.
Título da Revista
ISSN da Revista
Título de Volume
Editor
Resumo
Bioactivity of CoCrMo can be enhanced by incorporation of hydroxyapatite (HAP), but the poor mechanical properties of HAP is the most serious hindrance for its application on load-bearing implants, especially when exposed to corrosion and wear, simultaneously. Thus, the present work aims to investigate the tribocorrosion behaviour of CoCrMo−HAP composites. The tribocorrosion tests were performed under open circuit potential (OCP) in 8 g/l NaCl at 37 ºC by using a ball-on-plate tribometer. Results suggested that HAP particle addition increased the corrosion rate due to localized corrosion taking place on the pore sites near the matrix/reinforcement interface. However, under tribological action, composite samples presented relatively lower tendency to corrosion and lower coefficient of friction values.
Descrição
Palavras-chave
Corrosive degradation, Metal matrix composite, Tribochemistry
Como citar
Tribology International, v. 91, p. 221-227.