摘要: This paper presents the design and implementation of an integrated wireless microsystem platform that provides the possibility to support versatile implantable neural sensing devices in free laboratory rats. Inductive coupled coils with low dropout regulator design allows true long-term recording without limitation of battery capacity. A 16-channel analog front end chip located on the headstage is designed for high channel account neural signal conditioning with low current consumption and noise. Two types of implantable electrodes including grid electrode and 3D probe array are also presented for brain surface recording and 3D biopotential acquisition in the implanted target volume of tissue. The overall system consumes less than 20 mA with small form factor, 3.9 x 3.9 cm(2) mainboard and 1.8 x 3.4 cm(2) headstage, is packaged into a backpack for rats. Practical in vivo recordings including auditory response, brain resection tissue and PZT-induced seizures recording demonstrate the correct function of the proposed microsystem. Presented achievements addressed the aforementioned properties by combining MEMS neural sensors, low-power circuit designs and commercial chips into system-level integration.
URI: http://dx.doi.org/10.3390/s130404624
http://hdl.handle.net/11536/21943
ISSN: 1424-8220
DOI: 10.3390/s130404624
期刊: SENSORS
Volume: 13
Issue: 4
起始頁: 4624
結束頁: 4639
顯示於類別:期刊論文


文件中的檔案:

  1. 000318036400038.pdf

若為 zip 檔案,請下載檔案解壓縮後,用瀏覽器開啟資料夾中的 index.html 瀏覽全文。