Clinical Evaluation of Non-Contact Infrared Thermometers

Author/Creator ORCID

Date

2021-08-18

Department

Program

Citation of Original Publication

Sullivan, Stacey J.L. et al.; Clinical Evaluation of Non-Contact Infrared Thermometers; Scientific Reports, 18 August, 2021; https://doi.org/10.21203/rs.3.rs-802048/v1

Rights

This item is likely protected under Title 17 of the U.S. Copyright Law. Unless on a Creative Commons license, for uses protected by Copyright Law, contact the copyright holder or the author.
Attribution 4.0 International (CC BY 4.0)

Subjects

Abstract

Background: Non-contact infrared thermometers (NCITs) are being widely used during the COVID-19 pandemic as a temperature-measurement tool for screening and isolating patients in healthcare settings, travelers at ports of entry, and the general public. Methods: To understand the accuracy of NCITs, a clinical study was conducted with 1113 adult subjects using six different commercially available NCIT models. A total of 60 NCITs were tested with 10 units for each model. The NCIT-measured temperature was compared with the oral temperature obtained using a reference oral thermometer. Results: The mean difference between the reference thermometer and NCIT measurement (clinical bias) was different for each NCIT model. The clinical bias ranged from just under -0.9 °C (under-reporting) to just over 0.2 °C (over-reporting). The individual differences ranged from -3 °C to +2 °C in extreme cases, with the majority of the differences between -2 °C and +1 °C. Depending upon the NCIT model, 48% to 88% of the individual temperature measurements were outside the labeled accuracy stated by the manufacturers. The sensitivity of the NCIT models for detecting subject’s temperature above 38 °C ranged from 0 to 0.69. Conclusions: Overall, our results indicate that some NCIT devices may not be consistently accurate enough to determine if subject’s temperature exceeds a specific threshold of 38 °C. Model-to-model variability and individual model accuracy in the displayed temperature were found to be outside of acceptable limits. Accuracy and credibility of the NCITs should be thoroughly evaluated before using them as an effective screening tool.