Design of internal MRI coils using ultimate intrinsic SNR

Date
2009-09-27
Editor(s)
Advisor
Supervisor
Co-Advisor
Co-Supervisor
Instructor
Source Title
Magnetic Resonance Materials in Physics, Biology and Medicine
Print ISSN
0968-5243
Electronic ISSN
Publisher
Springer
Volume
22
Issue
4
Pages
221 - 228
Language
English
Journal Title
Journal ISSN
Volume Title
Series
Abstract

Object: Internal MRI coils have important applications in diagnostic and interventional studies. Since they can be placed very close to the region of interest in the body, they are favored over external coils in applications where high-resolution images are required. In this paper it is demonstrated that ultimate intrinsic SNR (UISNR) and the optimum coil sensitivity solutions can be used to make new coil designs with higher intrinsic SNR. Materials and methods: In this study, UISNR, which is the maximum attainable value of the intrinsic SNR, is used as a measure of performance and as a design criterion. As an example, a novel endorectal MRI coil is designed. The design is tested with phantom and patient studies. Results: An endorectal coil is built to demonstrate the effectiveness of the design strategy. ISNR of the endorectal coil approximates the UISNR to 72%. Conclusion: An internal coil design method that takes advantage of the UISNR and optimum coil sensitivity calculations was presented. This method can also be used to design better internal MRI coils for different applications.

Course
Other identifiers
Book Title
Citation
Published Version (Please cite this version)