Please use this identifier to cite or link to this item: http://hdl.handle.net/1942/12997
Title: Electrochemical sensor array for bioprocess monitoring
Authors: Bäcker, M.
DELLE, Lotta 
Poghossian, A.
Biselli, M.
Zang, W.
WAGNER, Patrick 
Schöning, M. J.
Issue Date: 2011
Publisher: PERGAMON-ELSEVIER SCIENCE LTD
Source: ELECTROCHIMICA ACTA, 56 (26), p. 9673-9678
Abstract: A silicon-based biosensor chip consisting of an array of amperometric enzyme sensors has been developed for monitoring the concentration of glutamate and glutamine in cell-culture fermentation processes. The glutamate sensor was constructed by immobilising glutamate oxidase, while the glutamine sensor by sequential coupling of glutaminase and glutamate oxidase. The enzymes were immobilised by means of cross-linking with glutaraldehyde on the surface of the patterned platinum thin-film electrodes. The developed biosensors have been electrochemically characterised in solutions with different concentrations of glutamate and glutamine in terms of sensitivity, response time, linear working range and lifetime. A high sensitivity of 96 nA/mM and 100 nA/mM was registered for the glutamate and glutamine sensors, respectively. The preliminary experiments in cell-culture medium have shown a good correlation between the glutamine and glutamate concentrations measured with the biosensor chip and the commercially available biochemistry analyser. The obtained results demonstrate the feasibility of the realised biosensor chip for monitoring the glutamine and glutamate concentrations in fermentation processes. (C) 2011 Elsevier Ltd. All rights reserved.
Notes: Schoning, MJ (reprint author); Aachen Univ Appl Sci, Inst Nano & Biotechnol INB, Campus Julich, Aachen, Germany. [Baecker, M.; Delle, L.; Poghossian, A.; Biselli, M.; Schoening, M. J.] Aachen Univ Appl Sci, Inst Nano & Biotechnol INB, Aachen, Germany. [Baecker, M.; Poghossian, A.; Schoening, M. J.] Res Ctr Julich GmbH, Peter Grunberg Inst PGI 8, Julich, Germany. [Zang, W.] HiTec Zang GmbH, Herzogenrath, Germany. [Wagner, P.] Hasselt Univ, Inst Mat Res, Diepenbeek, Belgium. schoening@fh-aachen.de
Keywords: Electrochemistry;Amperometric sensor array; Enzymatic biosensor; Bioprocess monitoring; Glutamine; Glutamate
Document URI: http://hdl.handle.net/1942/12997
ISSN: 0013-4686
e-ISSN: 1873-3859
DOI: 10.1016/j.electacta.2011.04.030
ISI #: 000297493300022
Category: A1
Type: Journal Contribution
Validations: ecoom 2012
Appears in Collections:Research publications

Files in This Item:
File Description SizeFormat 
backer1.pdf
  Restricted Access
Published version663.29 kBAdobe PDFView/Open    Request a copy
Show full item record

SCOPUSTM   
Citations

15
checked on Sep 2, 2020

WEB OF SCIENCETM
Citations

17
checked on Apr 14, 2024

Page view(s)

98
checked on Sep 7, 2022

Download(s)

84
checked on Sep 7, 2022

Google ScholarTM

Check

Altmetric


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.