NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
The development of insulated electrocardiogram electrodesAn integrated system was developed, consisting of an insulated electrode and an impedance transformer, which can be used for the acquisition of electrocardiographic data. The electrode consists of a thin layer of dielectric material deposited onto a silicon substrate. The impedance transformer is an operational amplifier used in the unity gain configuration. Both electrode and impedance transformer are contained in a plastic housing identical to that used with the NASA Apollo-type electrode. The lower cut off frequency of the electrode system is between 0.01 and 1.0 Hz, depending on the dielectric used and its thickness. Clinical quality electrocardiograms were obtained with these electrodes.
Document ID
19720014446
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Portnoy, W. M.
(Texas Technological Univ. Lubbock, TX, United States)
David, R. M.
(Texas Technological Univ. Lubbock, TX, United States)
Date Acquired
August 6, 2013
Publication Date
December 31, 1971
Subject Category
Biotechnology
Report/Patent Number
NASA-CR-115530
Accession Number
72N22096
Funding Number(s)
CONTRACT_GRANT: NAS9-11162
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.