Evaluation tests of platinum resistance thermometers for a cryogenic wind tunnel applicationThirty-one commercially designed platinum resistance thermometers were evaluated for applicability to stagnation temperature measurements between -190 C and +65 C in the Langley Research Center's National Transonic Facility. Evaluation tests included X-ray shadowgraphs, calibrations before and after aging, and time constant measurements. Two wire-wound low thermal mass probes of a conventional design were chosen as most suitable for this cryogenic wind tunnel application.