Flow Compensation in Non-Balanced SSFP

MPS-Authors
There are no MPG-Authors in the publication available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Santini, F., Bieri, O., & Scheffler, K. (2008). Flow Compensation in Non-Balanced SSFP. Poster presented at 16th Scientific Meeting and Exhibition of the International Society of Magnetic Resonance in Medicine (ISMRM 2008), Toronto, Canada.


Cite as: https://hdl.handle.net/21.11116/0000-0003-9334-C
Abstract
Flow compensation is a critical issue in non-balanced SSFP imaging because high first-order moments are achieved in the dephasing direction. Phase
accumulation leads to steady-state disruption and consequent blurring artifacts and signal loss, especially in presence of long-T2 tissues. In this work, a
theoretical explanation of the artifacts is given, showing how phase fluctuations influence the amplitude and the stability of the signal, and optimized flow
compensated SSFP-FID and SSFP-echo sequences are presented. Finally, In vivo application of the sequence is demonstrated.