English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Paper

Spectral Intensity Bioaerosol Sensor (SIBS): A new Instrument for Spectrally Resolved Fluorescence Detection of Single Particles in Real-Time

MPS-Authors
/persons/resource/persons203107

Könemann,  Tobias
Multiphase Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons101066

Klimach,  Thomas
Particle Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons127609

Walter,  David
Atmospheric Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons100944

Fröhlich-Nowoisky,  Janine
Multiphase Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons101295

Su,  Hang
Multiphase Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons101189

Pöschl,  Ulrich
Multiphase Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

/persons/resource/persons104597

Pöhlker,  Christopher
Multiphase Chemistry, Max Planck Institute for Chemistry, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Könemann, T., Savage, N., Klimach, T., Walter, D., Fröhlich-Nowoisky, J., Su, H., et al. (2018). Spectral Intensity Bioaerosol Sensor (SIBS): A new Instrument for Spectrally Resolved Fluorescence Detection of Single Particles in Real-Time. Atmospheric Measurement Techniques Discussions, 11.


Cite as: https://hdl.handle.net/21.11116/0000-0003-0F76-9
Abstract
Primary biological aerosol particles (PBAP) in the atmosphere are highly relevant for the Earth system, climate, and public health. The analysis of PBAP, however, remains challenging due to their high diversity and large spatiotemporal variability. For real-time PBAP analysis, light-induced fluorescence (LIF) instruments have been developed and widely used in laboratory and ambient studies. The interpretation of fluorescence data from these instruments, however, is often limited by a lack of spectroscopic information. This study introduces a new instrument – the Spectral Intensity Bioaerosol Sensor (SIBS) – that resolves fluorescence spectra for single particles and, thus, promises to expand the scope of fluorescent PBAP quantification and classification.

The SIBS shares key design components with the latest versions of the Wideband Integrated Bioaerosol Sensor (WIBS) and the findings presented here are also relevant for the widely deployed WIBS-4A and WIBS-NEO as well as other LIF instruments. The key features of the SIBS and findings of this study can be summarized as follows:

– Particle sizing yields reproducible linear responses for particles in the range of 300 nm to 20 µm. The lower sizing limit is significantly smaller than for earlier commercial LIF instruments (e.g., WIBS-4A and the Ultraviolet Aerodynamic Particle Sizer (UV-APS)), expanding the analytical scope into the accumulation mode size range.
– Fluorescence spectra are recorded for two excitation wavelengths (λex = 285 and 370 nm) and a wide range of emission wavelengths (λmean = 302–721 nm) with a resolution of 16 detection channels, which is higher than for most other commercially available LIF bioaerosol sensors.

– Fluorescence spectra obtained for 16 reference compounds confirm that the SIBS provides sufficient spectral resolution to distinguish major modes of molecular fluorescence. For example, the SIBS resolves the spectral difference between bacteriochlorophyll and chlorophyll a/b.

– A spectral correction of the instrument-specific detector response is essential to use the full fluorescence emission range.

– Asymmetry factor (AF) data were assessed and were found to provide only limited analytical information.

– In test measurements with ambient air, the SIBS worked reliably and yielded characteristically different spectra for single particles in the coarse mode with an overall fluorescent particle fraction of ~ 4 % (3σ threshold), which is consistent with earlier studies in comparable environments.