Analysis of EPI phase correction with low flip‐angle excitation to reduce the required minimum TE: Application to whole‐brain, submillimeter‐resolution fMRI at 3 T

 ;

2020
Wiley-Liss New York, NY [u.a.]

Magnetic resonance in medicine 84(3), 1416 - 1429 () [10.1002/mrm.28218]

This record in other databases:      

Please use a persistent id in citations:   doi:

Abstract: Echo planar imaging is used widely for its imaging speed. However, its applications often suffer from ghost artifacts. In the community, an approach using three navigator echoes is used commonly for the artifact correction. Although this scheme is effective, as the matrix size increases for high‐resolution imaging, the navigator echoes can contribute significantly to increasing the “required minimum TE.” To overcome this issue, this work proposes the use of an alternative navigator echo scheme called the “TR‐external” scheme.The TR‐external scheme reduces the required minimum TE by allocating an additional excitation loop for the navigator echoes before every main excitation loop. In this work, a detailed analysis on the TR‐external scheme was performed to assess its performance in comparison to the standard scheme. Visual fMRI was performed to check the feasibility of using the TR‐external scheme for detecting functional signals.The performance of the TR‐external scheme was comparable with that of the standard scheme in terms of the SNR, elimination of ghost artifacts, and the BOLD detection. For a given matrix size (288 × 288), the TR‐external scheme allowed a substantially shorter TE (5.94 ms) compared with the standard scheme, which resulted in a higher SNR. Furthermore, this feature enabled the submillimeter‐resolution (0.73 × 0.73 mm2) fMRI measurement with a favorable TE (35 ms) at 3 T. The fMRI results revealed that activated voxels are well localized along the cortical ribbon.A TR‐external scheme for EPI phase correction was implemented at 3 T. Its feasibility for submillimeter‐resolution fMRI was successfully demonstrated.

Classification:

Contributing Institute(s):
  1. Jara-Institut Quantum Information (INM-11)
  2. Physik der Medizinischen Bildgebung (INM-4)
  3. JARA-BRAIN (JARA-BRAIN)
Research Program(s):
  1. 573 - Neuroimaging (POF3-573) (POF3-573)

Appears in the scientific report 2020
Database coverage:
Medline ; Creative Commons Attribution CC BY 4.0 ; OpenAccess ; BIOSIS Previews ; Clarivate Analytics Master Journal List ; Current Contents - Clinical Medicine ; Current Contents - Life Sciences ; IF < 5 ; JCR ; NationallizenzNationallizenz ; SCOPUS ; Science Citation Index ; Science Citation Index Expanded ; Web of Science Core Collection
Click to display QR Code for this record

The record appears in these collections:
Document types > Articles > Journal Article
JARA > JARA > JARA-JARA\-BRAIN
Institute Collections > INM > INM-11
Institute Collections > INM > INM-4
Workflow collections > Public records
Workflow collections > Publication Charges
Publications database
Open Access

 Record created 2020-03-05, last modified 2022-09-30