Commercial 18 carat (ct) gold alloys along with pure coinage metals have been studied with Knudsen effusion mass spectrometry. Isotopic fractionation in vapour phase and the enthalpies of vaporization were estimated for Au, Ag and Cu samples. The assaying of the gold content was carried out by means of calibration with respect to standard reference alloys measured with energy dispersive X-ray spectroscopy. The accuracy of the gold determination resulted of about 1.5 wt parts per thousand in the ternary Au-Ag-Cu alloy.

Gold assay with Knudsen effusion mass spectrometry / U. Bardi; F. Niccolai; M. Tosti; A. Tolstogouzov. - In: INTERNATIONAL JOURNAL OF MASS SPECTROMETRY. - ISSN 1387-3806. - STAMPA. - 273:(2008), pp. 138-144. [10.1016/j.ijms.2008.03.013]

Gold assay with Knudsen effusion mass spectrometry

BARDI, UGO;
2008

Abstract

Commercial 18 carat (ct) gold alloys along with pure coinage metals have been studied with Knudsen effusion mass spectrometry. Isotopic fractionation in vapour phase and the enthalpies of vaporization were estimated for Au, Ag and Cu samples. The assaying of the gold content was carried out by means of calibration with respect to standard reference alloys measured with energy dispersive X-ray spectroscopy. The accuracy of the gold determination resulted of about 1.5 wt parts per thousand in the ternary Au-Ag-Cu alloy.
2008
273
138
144
U. Bardi; F. Niccolai; M. Tosti; A. Tolstogouzov
File in questo prodotto:
File Dimensione Formato  
Gold assay with Knudsen effusion.pdf

accesso aperto

Tipologia: Versione finale referata (Postprint, Accepted manuscript)
Licenza: Open Access
Dimensione 624.96 kB
Formato Adobe PDF
624.96 kB Adobe PDF

I documenti in FLORE sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificatore per citare o creare un link a questa risorsa: https://hdl.handle.net/2158/778983
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 8
social impact