タイトル: | Field-ionization electron detector at low temperature of 10 mK range |
著者: | Shibata, M Tada, M Kishimoto, Y https://orcid.org/0000-0002-3049-9063 (unconfirmed) Kominato, K Haseyama, T Ogawa, I Matsuki, S Yamada, S Funahashi, H Yamamoto, K |
発行日: | Jul-2003 |
出版者: | American Institute of Physics |
誌名: | REVIEW OF SCIENTIFIC INSTRUMENTS |
巻: | 74 |
号: | 7 |
開始ページ: | 3317 |
終了ページ: | 3323 |
抄録: | Selective field-ionization electron detector at low temperature of 10 mK range was developed with a channel electron multiplier. The field-ionization electrode system is attached to the bottom plate of the mixing chamber of a dilution refrigerator and ionized electrons are transported to a channel electron multiplier at the 1 K temperature pumping stage through a series of ring focusing electrodes. The channel electron multiplier is heated up to more than 20 K with a heating coil to maintain its long-term operational conditions, while keeping its environment temperature to 1 K. Rydberg states in Rb with 80≲n≲150 were successfully field ionized and detected with this system at the long-term operating temperature of 12–15 mK. |
著作権等: | Copyright 2003 American Institute of Physics. This article may be downloaded for personal use only. Any other use requires prior permission of the author and the American Institute of Physics. |
URI: | http://hdl.handle.net/2433/39807 |
DOI(出版社版): | 10.1063/1.1582392 |
出現コレクション: | 学術雑誌掲載論文等 |