Efficacy of UV-C disinfection in hyperbaric chambers.

Abstract

Ultraviolet C (UV-C) light reduces contamination on high-touch clinical surfaces. We assessed the efficacy of 2 UV-C devices at eradicating important clinical pathogens in hyperbaric chambers. Both devices were similarly efficacious against MRSA but differed significantly against C. difficile. Additionally, direct UV-C exposure was more efficacious against both species than indirect exposure.

Department

Description

Provenance

Subjects

Citation

Published Version (Please cite this version)

10.1017/ice.2020.248

Publication Info

Warren, Bobby Glenn, Jason Masker, Gregory Brown, Isabella Gamez, Becky Smith, Deverick J Anderson and Nicholas Turner (2020). Efficacy of UV-C disinfection in hyperbaric chambers. Infection control and hospital epidemiology, 41(1). pp. 1–4. 10.1017/ice.2020.248 Retrieved from https://hdl.handle.net/10161/21097.

This is constructed from limited available data and may be imprecise. To cite this article, please review & use the official citation provided by the journal.

Scholars@Duke

Anderson

Deverick John Anderson

Professor of Medicine

Hospital epidemiology, infection control, antibiotic stewardship, multidrug-resistant organisms, device-related infections, surgical site infections, catheter-associated bloodstream infections, cost of infections, infections in community hospitals

Turner

Nicholas Turner

Assistant Professor of Medicine

Unless otherwise indicated, scholarly articles published by Duke faculty members are made available here with a CC-BY-NC (Creative Commons Attribution Non-Commercial) license, as enabled by the Duke Open Access Policy. If you wish to use the materials in ways not already permitted under CC-BY-NC, please consult the copyright owner. Other materials are made available here through the author’s grant of a non-exclusive license to make their work openly accessible.