Compartir
Título
Rapid and reliable analysis of underivatized amino acids in urine using tandem mass spectrometry
Autor(es)
Palabras clave
Stand-alone tandem mass spectrometry
Guard column
Tandem mass spectrometry
LC-MS
Amino acids
Urine samples
Clasificación UNESCO
2301 Química Analítica
Fecha de publicación
2022
Editor
Elsevier
Citación
Fernández-del-Campo-García, M. T., Casas-Ferreira, A. M., Rodríguez-Gonzalo, E., Moreno-Cordero, B., & Pérez-Pavón, J. L. (2022). Rapid and reliable analysis of underivatized amino acids in urine using tandem mass spectrometry. Microchemical Journal, 172, 106914. https://doi.org/10.1016/j.microc.2021.106914
Resumen
[EN] Nowadays there is a great interest in the development of fast and reliable methods for the determination of amino
acids (AAs) in biological samples due to their biological importance. In the present work, a method based on the
use of a guard column (gC) prior to tandem mass spectrometry (gC-MS/MS) is proposed for the determination of
proteinogenic AAs in urine. Heptafluorobutyric acid (HFBA) is used as ion pairing reagent. Comparison of the gCMS/MS method versus tandem mass spectrometry (MS/MS) in stand-alone mode showed improved sensitivity
and peak shape, and solved some problems related to interfering compounds, with a total analysis time of 2.8
min. All the proteinogenic AAs were adequately determined using the gC-MS/MS method, except glutamic acid
(Glu).
To confirm quantitative results obtained with gC-MS/MS for individual AAs, an ion-pair liquid chromatography tandem mass spectrometry method (LC-MS/MS) has also been developed. Both methods (gC-MS/MS and
LC-MS/MS) were validated using synthetic urine. For the gC-MS/MS method, LODs and LOQs values were found
to be between 0.004 and 0.425 mg/L and 0.01 and 1.40 mg/L, respectively. Aspartic acid (Asp) showed the
highest LOD and LOQ values (3 mg/L and 9 mg/L, respectively). A one-point standard addition method and
internal standard normalization were used for the quantification because matrix effects were observed. Lalanine-1-13C (Ala-13C) and L-leucine-1-13C (Leu-13C) were used as isotopically labeled internal standards.
To demonstrate the applicability of the gC-MS/MS method in the reliable determination of AAs in real samples, urine from eighteen healthy volunteers were analyzed using both gC-MS/MS and LC-MS/MS methods.
Similar quantitative results were obtained for individual AAs with both of them. In addition, possible differences
in AAs concentrations related to sex were checked, but the results did not show significant differences for the
evaluated compounds.
URI
ISSN
0026-265X
DOI
10.1016/j.microc.2021.106914
Versión del editor
Aparece en las colecciones
- DQANB. Artículos [77]
Patrocinador
Publicación en abierto financiada por la Universidad de Salamanca como participante en el Acuerdo Transformativo CRUE-CSIC con Elsevier, 2021-2024
Ficheros en el ítem
Tamaño:
324.9Kb
Formato:
Adobe PDF
Descripción:
Material suplementario