Properties of focusing polycapillary utilized in .sup.57./sup.Fe Mössbauer spectroscopy

  1. 1.
    0561959 - FZÚ 2023 RIV GB eng J - Článek v odborném periodiku
    Kočiščák, J. - Novák, P. - Chmelíčková, Hana - Procházka, V.
    Properties of focusing polycapillary utilized in 57Fe Mössbauer spectroscopy.
    Measurement. Roč. 192, Mar (2022), č. článku 110842. ISSN 0263-2241. E-ISSN 1873-412X
    Institucionální podpora: RVO:68378271
    Klíčová slova: Mössbauer spectrometer * MCX * polycapillary optics
    Obor OECD: Nuclear physics
    Impakt faktor: 5.6, rok: 2022
    Způsob publikování: Omezený přístup
    https://doi.org/10.1016/j.measurement.2022.110842

    A focusing polycapillary was incorporated in the Mossbauer spectrometer to achieve Mossbauer spectrum from the point with the diameter of 150 mu m. Since such an apparatus is rather unique and so far it has not been described, a detailed views into the problematics of using the focusing polycapillary with Co-57 source, and into the problematics of the scanning transmission Mossbauer microscope alignment procedure are presented. This article presents the properties of the focusing polycapillary and its performance for Fe-57 14.4 keV Mossbauer radiation using Co-57 radioactive source. The performance of the polycapillary is demonstrated with the measurement of the stainless steel. The obtained spectrum exhibited a similar signal to noise ratio as the Mossbauer spectrum acquired by using a pinhole with the diameter of 200 mu m. The usage of the polycapillary optics could be a promising way in Mossbauer spectroscopy, e.g. for measurements of tiny samples in an anvil cell.

    Trvalý link: https://hdl.handle.net/11104/0334389

     
     
Počet záznamů: 1  

  Tyto stránky využívají soubory cookies, které usnadňují jejich prohlížení. Další informace o tom jak používáme cookies.