Development of a Carbon Mesh Supported Thin Film Microextraction Membrane As a Means to Lower the Detection Limits of Benchtop and Portable GC/MS Instrumentation

Download
2016-02-02
Grandy, Jonathan J.
Boyacı, Ezel
Pawliszyn, Janusz
In this work, a durable and easy to handle thin film microextraction (TFME) device is reported. The membrane is comprised of poly(divinylbenzene) (DVB) resin particles suspended in a high-density polydimethylsiloxane (PDMS) glue, which is spread onto a carbon fiber mesh. The currently presented membrane was shown to exhibit a substantially lesser amount of siloxane bleed during thermal desorption, while providing a statistically similar extraction efficiency toward a broad spectrum of analytes varying in polarity when compared to an unsupported DVB/PDMS membrane of similar shape and size which was prepared with previously published methods. With the use of hand-portable GC-TMS instrumentation, membranes cut with dimensions 40 mm long by 4.85 mm wide and 40 +/- 5 mu m thick (per side) were shown to extract 21.2, 19.8, 18.5, 18,4, 26.8, and 23.7 times the amount of 2,4 dichlorophenol, 2,4,6 trichlorophenol, phorate D10, fonofos, chloropyrifos, and parathion, respectively, within 15 min from a 10 ppb aqueous solution as compared to a 65 mu m DVB/PDMS solid phase microextraction (SPME) fiber. A portable high volume desorption module prototype was also evaluated and shown to be appropriate for the desorption of analytes with a volatility equal to or lesser than benzene when employed in conjunction with TFME membranes. Indeed, the coupling of these TFME devices to hand-portable gas chromatography toroidial ion trap mass spectrometry (GC-TMS) instrumentation was shown to push detection limits for these pesticides down to the hundreds of ppt levels, nearing that which can be achieved with benchtop instrumentation. Where these membranes can also be coupled to benchtop instrumentation it is reasonable to assume that detection limits could be pushed down even further. As a final proof of the concept, the first ever, entirely on-site TFME-GC-TMS analysis was performed at a construction impacted lake. Results had indicated the presence of contaminants such as toluene, ethylbenzene, xylene, 2,2,4-trimethyl-1,3-pentanediol diisobutyrate, and tris(1-chloro-2-propyl)phosphate, which stood out from other naturally occurring compounds detected.
ANALYTICAL CHEMISTRY

Suggestions

New Generation of Solid-Phase Microextraction Coatings for Complementary Separation Approaches: A Step toward Comprehensive Metabolomics and Multiresidue Analyses in Complex Matrices
Gionfriddo, Emanuela; Boyacı, Ezel; Pawliszyn, Janusz (American Chemical Society (ACS), 2017-04-04)
In this work, a new generation of solid-phase microextraction (SPME) coatings based on polytetrafluoro-ethylene amorphous fluoroplastics (PTFE AF 2400) as a particle binder is presented. The developed coating was tested for thermal and solvent-assisted desorption, demonstrating its compatibility with both gas- and liquid-chromatographic platforms. The incorporation of hydrophilic-lipophilic balance (HLB) adsorptive particles provided optimal extraction coverage for analytes bearing a broad range of hydropho...
Use of a simple transient extension chamber with ETV-ICPMS: quantitative analysis and matrix effects
Ertaş, Gülay (Royal Society of Chemistry (RSC), 2003-01-01)
The transient extension (TEx) chamber was developed to provide a simple means of lengthening an electrothermal vaporizer (ETV) signal for the purpose of obtaining a full mass scan from a single ETV firing with inductively coupled plasma mass spectrometry (ICPMS) detection. The TEx chamber was used for quantitative analysis of natural water (NIST SRM 1640). Quantitative analysis was done for Co, Be, Pb, Sb and Cd. Detection limits for the five elements tested with the TEx chamber were in the 1-10 mug L-1 ran...
Synthesis of colloidal silver particles with different sizes by seeding approach for surface enhanced raman scattering (sers) studies
Sancı, Rukiye; Volkan, Mürvet; Department of Chemistry (2009)
In this study, silver nanorods and nanospheroids were prepared both in aqueous solution and on the surface of glass slides through seed-mediated growth approach at room temperature and used as a surface enhanced Raman scattering (SERS) substrate. The synthesis of metallic nanorods was started with the production of silver nanospheres as seed utilizing sodium borohydride and trisodium citrate as reducing and capping agents, respectively. These seeds were then added to a growth solution containing additional ...
Isochronal hydrogenation of textured magnesium/palladium thin films
Özgit, Çağla; Öztürk, Tayfur; Department of Metallurgical and Materials Engineering (2009)
Pure and palladium‐covered 350 nm thick magnesium thin films were deposited on glass substrates via thermal evaporation. In the as‐deposited state, films were highly textured with Mg (001) parallel to the substrate. Hydrogen loading experiments were carried out in two different conditions; namely isothermal and isochronal. Hydrogenation behaviors of the thin films were followed by twopoint probe electrical resistance and optical transmittance measurements, as well as x‐ray diffraction studies. Isothermal hy...
Immobilization studies utilizing solid supports for the determination of fructose by dansylaminophenyl boronic acid (DAPB acid) and chromate by diphenylcarbazide (DPC)
Bulut, Mukadder; Volkan, Mürvet; Department of Chemistry (2006)
Immobilization of fluorescent chemosensors and chromogenic reagents on solid supports for developing optical sensors result in improved analytical performance characteristics such as continuous read-out, increased sensitivity, lower reagent consumption and possibility of using the sensor in solvents where the free molecule displays low solubility. The aim of this study is to immobilize dansylaminophenyl boronic acid (DAPB acid) and diphenylcarbazide (DPC) into various solid supports for the determination of...
Citation Formats
J. J. Grandy, E. Boyacı, and J. Pawliszyn, “Development of a Carbon Mesh Supported Thin Film Microextraction Membrane As a Means to Lower the Detection Limits of Benchtop and Portable GC/MS Instrumentation,” ANALYTICAL CHEMISTRY, pp. 1760–1767, 2016, Accessed: 00, 2020. [Online]. Available: https://hdl.handle.net/11511/39901.