English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT

Released

Journal Article

Rapid and motion-robust volume coverage using cross-sectional real-time MRI.

MPS-Authors
/persons/resource/persons15968

Voit,  D.
Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society;

/persons/resource/persons205125

Kalentev,  O.
Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society;

/persons/resource/persons205128

van Zalk,  M.
Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society;

/persons/resource/persons59192

Joseph,  A. A.
Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society;

/persons/resource/persons15082

Frahm,  J.
Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)

3181841.pdf
(Preprint), 629KB

Supplementary Material (public)

3181841_Suppl_1.pdf
(Supplementary material), 2MB

3181841_Suppl_2.pdf
(Supplementary material), 16MB

Citation

Voit, D., Kalentev, O., van Zalk, M., Joseph, A. A., & Frahm, J. (2019). Rapid and motion-robust volume coverage using cross-sectional real-time MRI. Magnetic Resonance in Medicine, (in press). doi:10.1002/mrm.28029.


Cite as: https://hdl.handle.net/21.11116/0000-0005-5929-A
Abstract
PURPOSE:

To develop a rapid and motion-robust technique for volumetric MRI, which is based on cross-sectional real-time MRI acquisitions with automatic advancement of the slice position.
METHODS:

Real-time MRI with a frame-by-frame moving cross-section is performed with use of highly undersampled radial gradient-echo sequences offering spin density, T1 , or T2 /T1 contrast. Joint reconstructions of serial images and coil sensitivity maps from spatially overlapping sections are accomplished by nonlinear inversion with regularization to the preceding section-formally identical to dynamic real-time MRI. Shifting each frame by 20% to 25% of the section thickness ensures 75% to 80% overlap of successive sections. Acquisition times of 40 to 67 ms allow for rates of 15 to 25 sections per second, while volumes are defined by the number of cross-sections times the section shift.
RESULTS:

Preliminary realizations at 3T comprise studies of the human brain, carotid arteries, liver, and prostate. Typically, coverage of a 90- to 180-mm volume at 0.8- to 1.2-mm in-plane resolution, 4- to 6-mm section thickness, and 0.8- to 1.5-mm section shift is accomplished within total measuring times of 4 to 6 seconds and a section speed of 15 to 37.5 mm per second. However, spatiotemporal resolution, contrast including options such as fat saturation and total measuring time are highly variable and may be adjusted to clinical needs. Promising volumetric applications range from fetal MRI to dynamic contrast-enhanced MRI.
CONCLUSION:

The proposed method allows for rapid and motion-robust volume coverage in a variety of imaging scenarios.