TIDE (transition into driven equilibrium)-Sequences for Brain Imaging with Improved Signal and Contrast.Behaviour

MPS-Authors
There are no MPG-Authors in the publication available
External Resource
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Hennig, J., & Scheffler, K. (2003). TIDE (transition into driven equilibrium)-Sequences for Brain Imaging with Improved Signal and Contrast.Behaviour. Poster presented at 11th Scientific Meeting of the International Society of Magnetic Resonance in Medicine (ISMRM 2003), Toronto, ON, Canada.


Cite as: https://hdl.handle.net/21.11116/0000-0005-7799-9
Abstract
The TIDE-sequence is based on a steady state free precession (SSFP) experiment, where variable flip angles are used to bring the spin system in and out of the signal steady state. The purpose of this paper is to demonstrate new implementations of TIDE for 2D- and 3D-applications, where the high S/N and the possibility to start data acquisition without delay are used to create images with contrast so far unattainable with SSFP-sequences. Especially MP-TIDE which is based on similar principles as MR-RAGE is demonstrated to produce images with very high gray-white matter contrast.