日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細


公開

学術論文

A beetle-type, variable temperature scanning tunneling microscope for video-rate imaging

MPS-Authors
/persons/resource/persons39377

Platschkowski,  Viktor
Molecular Physics, Fritz Haber Institute, Max Planck Society;

External Resource
There are no locators available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)
公開されているフルテキストはありません
付随資料 (公開)
There is no public supplementary material available
引用

Henß, A.-K., Wiechers, J., Schuster, R., Platschkowski, V., & Wintterlin, J. (2020). A beetle-type, variable temperature scanning tunneling microscope for video-rate imaging. Japanese Journal of Applied Physics, 59:. doi:10.35848/1347-4065/ab8680.


引用: https://hdl.handle.net/21.11116/0000-0006-6CF7-B
要旨
A setup is described that combines video-rate imaging with temperature variability for a beetle-type scanning tunneling microscope (STM). The STM is operated in the constant height mode. Sinusoidal driving voltages are used for the scan lines; the resulting image distortions are corrected by a straightforward mathematical operation. A phase shifter is used to compensate the time lag between the driving voltage and the tip motion. A preamplifier with a bandwidth of 300 kHz and an rms noise of ~0.1 nA has been developed. Software has been written for data acquisition rates of up to 2 MHz and for saving movies of more than 3700 STM images of 200 × 200 pixels. Frame rates between 10 and 50 frames s−1 have been achieved, where the highest line rates are above the lowest resonance frequency of the STM. The accessible temperature range of 50–500 K is not impaired by the video mode.