Article (Scientific journals)
Validation of the quantitative determination of tetrahydrocannabinol and its two major metabolites in plasma by UHPLC/MS-MS according to the total error approach
DUBOIS, Nathalie; Paccau, A.; DE BACKER, Benjamin et al.
2012In Journal of Analytical Toxicology, 36 (1), p. 25-29
Peer Reviewed verified by ORBi
 

Files


Full Text
Validation of the Quantitative Determination of Tetrahydrocannabinol and Its Two Major Metabolites in Plasma.pdf
Publisher postprint (266.62 kB)
Download

http://jat.oxfordjournals.org/


All documents in ORBi are protected by a user license.

Send to



Details



Abstract :
[en] In Belgium, driving under the influence (DUI) of cannabis is prohibited and has severe legal consequences for the driver if the blood plasma concentration of Δ9-tetrahydrocannabinol (THC) exceeds 1 µg/L. A method to quantify low concentrations of THC and its hydroxylated (THC-OH) and carboxylated (THC-COOH) metabolites in plasma was developed for DUI but also for other applications. Ultra-high-performance liquid chromatography coupled to mass spectrometry seems to be a very convenient method to combine fast chromatographic separation and good sensitivity. The method was validated according to total error approach. Chromatographic separation was achieved in a 3-min total run time. The limits of quantitation were lower or equal to 1 µg/L for all compounds. The linearity of the method was acceptable in the validated range of concentrations (from 0.5 to 50 µg/L for THC, from 0.9 to 50 µg/L for THC-OH and from 1.1 to 100 µg/L for THC-COOH). The biases were lower than 13%, and the relative standard deviations for repeatability and intermediate precision did not exceed 15%. Lower and upper β-expectation tolerance limits did not exceed the acceptance limits of 20% for concentrations higher than 2 µg/L for THC and THC-OH and higher than 4 µg/L for THC-COOH. The acceptance limits were 30% for THC and THC-OH concentrations lower than 2 µg/L and for THC-COOH concentrations lower than 4 µg/L.
Disciplines :
Pharmacy, pharmacology & toxicology
Author, co-author :
DUBOIS, Nathalie ;  Centre Hospitalier Universitaire de Liège - CHU > Toxicologie clinique
Paccau, A.
DE BACKER, Benjamin ;  Centre Hospitalier Universitaire de Liège - CHU > Frais communs Biologie Clinique
Charlier, Corinne  ;  Université de Liège - ULiège > Département de pharmacie > Chimie toxicologique
Language :
English
Title :
Validation of the quantitative determination of tetrahydrocannabinol and its two major metabolites in plasma by UHPLC/MS-MS according to the total error approach
Publication date :
2012
Journal title :
Journal of Analytical Toxicology
ISSN :
0146-4760
eISSN :
1945-2403
Publisher :
Preston Publications Inc, Niles, United States - Illinois
Volume :
36
Issue :
1
Pages :
25-29
Peer reviewed :
Peer Reviewed verified by ORBi
Available on ORBi :
since 15 February 2012

Statistics


Number of views
242 (22 by ULiège)
Number of downloads
551 (7 by ULiège)

Scopus citations®
 
17
Scopus citations®
without self-citations
7
OpenCitations
 
19

Bibliography


Similar publications