Influence of a Fluoridated Medium with Different pHs on Commercially Pure Titanium-Based Implants

Nenhuma Miniatura disponível

Data

2009-02-01

Título da Revista

ISSN da Revista

Título de Volume

Editor

Wiley-Blackwell

Resumo

Purpose: The objective of this study was to assess the influence of a fluoride medium with different pHs on the corrosion resistance of three commercially pure titanium-based dental implant commercial brands, under scanning electron microscopy (SEM) and EDS.Materials and Methods: Forty-two dental implants, from three commercial brands, were used. Five years of regular use of mouth rinsing, with NaF 1500 ppm content and two different pHs, were simulated by immersing the specimens into that medium for 184 hours.Results: SEM and EDS analyses demonstrated no evidence of corrosion on the specimens' surfaces after being submitted to fluoride ions or incorporation of fluoride ions to the set surface.Conclusion: It was possible to conclude that both the fluoride concentration and the pH of the solutions did not exert any influence upon implant corrosion resistance.

Descrição

Palavras-chave

Corrosion, fluoride, dental implants, titanium

Como citar

Journal of Prosthodontics-implant Esthetic and Reconstructive Dentistry. Malden: Wiley-blackwell, v. 18, n. 2, p. 130-134, 2009.