A simple low-SAR technique for chemical-shift selection with high-field spin-echo imaging

MPS-Authors
/persons/resource/persons19748

Ivanov,  Dimo
Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

/persons/resource/persons19963

Schäfer,  Andreas
Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

/persons/resource/persons20026

Streicher,  Markus
Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

/persons/resource/persons19702

Heidemann,  Robin M.
Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

/persons/resource/persons20053

Trampel,  Robert
Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

/persons/resource/persons20055

Turner,  Robert
Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Supplementary Material (public)
There is no public supplementary material available
Citation

Ivanov, D., Schäfer, A., Streicher, M., Heidemann, R. M., Trampel, R., & Turner, R. (2010). A simple low-SAR technique for chemical-shift selection with high-field spin-echo imaging. Magnetic Resonance in Medicine, 64(2), 319-326. doi:10.1002/mrm.22518.


Cite as: https://hdl.handle.net/11858/00-001M-0000-0010-9E19-6
Abstract
We have discovered a simple and highly robust method for removal of chemical shift artifact in spin-echo MR images, which simultaneously decreases the radiofrequency power deposition (specific absorption rate). The method is demonstrated in spin-echo echo-planar imaging brain images acquired at 7 T, with complete suppression of scalp fat signal. When excitation and refocusing pulses are sufficiently different in duration, and thus also different in the amplitude of their slice-select gradients, a spatial mismatch is produced between the fat slices excited and refocused, with no overlap. Because no additional radiofrequency pulse is used to suppress fat, the specific absorption rate is significantly reduced compared with conventional approaches. This enables greater volume coverage per unit time, well suited for functional and diffusion studies using spin-echo echo-planar imaging. Moreover, the method can be generally applied to any sequence involving slice-selective excitation and at least one slice-selective refocusing pulse at high magnetic field strengths. The method is more efficient than gradient reversal methods and more robust against inhomogeneities of the static (polarizing) field (B0).