Wavelength calibration of optical Multichannel detectors in combination with single- and double-grating monochromators.

MPS-Authors
/persons/resource/persons15581

Nickel,  B.
Department of Spectroscopy and Photochemical Kinetics, MPI for biophysical chemistry, Max Planck Society;

External Resource
There are no locators available
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
フルテキスト (公開)

2376531.pdf
(出版社版), 506KB

付随資料 (公開)
There is no public supplementary material available
引用

Lindrum, M., & Nickel, B. (1989). Wavelength calibration of optical Multichannel detectors in combination with single- and double-grating monochromators. Applied Spectoscopy, 43(8), 1427-1431. doi:10.1366/0003702894204209.


引用: https://hdl.handle.net/11858/00-001M-0000-002C-27F2-E
要旨
For the measurement of a spectrum with an optical multichannel detector (OMCD), the relation N = N(λ) or λ = λ(N) between the channel number N and the wavelength λ must be known. For an ideal OMCD, N(λ) and λ(N) are completely defined by the physical parameters of the monochromator and the channel spacing of the OMCD. Series expansions N = N0 + ΣkAk(λ – λ0)k and λ = λ0 + ΣkBk(N – N0)k are given for single-grating monochromators (k = 1, 2, 3) and double-grating monochromators (k = 1, 2). The validity of the series expansions has been checked experimentally and numerically for a double monochromator.