Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT

Freigegeben

Zeitschriftenartikel

The extracellular patch clamp: A method for resolving currents through individual open channels in biological membranes

MPG-Autoren
/persons/resource/persons15570

Neher,  Erwin       
Department of Membrane Biophysics, MPI for biophysical chemistry, Max Planck Society;

/persons/resource/persons15741

Sakmann,  Bert
Department of Membrane Biophysics, MPI for biophysical chemistry, Max Planck Society;

Externe Ressourcen
Es sind keine externen Ressourcen hinterlegt
Volltexte (beschränkter Zugriff)
Für Ihren IP-Bereich sind aktuell keine Volltexte freigegeben.
Volltexte (frei zugänglich)
Es sind keine frei zugänglichen Volltexte in PuRe verfügbar
Ergänzendes Material (frei zugänglich)
Es sind keine frei zugänglichen Ergänzenden Materialien verfügbar
Zitation

Neher, E., Sakmann, B., & Steinbach, J. (1978). The extracellular patch clamp: A method for resolving currents through individual open channels in biological membranes. Pflügers Archiv: European Journal of Physiology, 375(2), 219-228. doi:10.1007/BF00584247.


Zitierlink: https://hdl.handle.net/21.11116/0000-000C-0EC0-C
Zusammenfassung
The current contributions of individual ionic channels can be measured by electrically isolating a small patch of membrane. To do this, the tip of a small pipette is brought into close contact with an enzymatically cleaned membrane of a hypersensitive amphibian or mammalian muscle fiber. Current flowing through the pipette is measured. If the pipette contains cholinergic agonist at μ-molar concentrations, square pulse current waveforms can be observed which represent the activation of individual acetylcholine-receptor channels. The square pulses have amplitudes of 1 to 3 pA and durations of 10–100 ms.

In order to obtain the necessary resolution, a delicate compromise had to be found between different experimental parameters. Pipettes with 1–3 μm internal diameter and a steep final taper had to be used, extensive enzyme treatment was necessary, and conditions had to be found in which channels open at a relatively low frequency.